ANESTHESIA MACHINE
       DR. PRIYANKA
ANESTHESIA MACHINE


• The function of anesthesia
  machine is preparation of
  anesthesia gas mixture of pure
  and precisely known yet
  variable composition at
  controlled & known pressure.
MACHINE STANDARDS

• 1979 – ANSI
• 1988 – ASTM

• Defined
  design, performance, safety
  requirements of anesthesia
  machine.
ANESTHESIA MACHINE
ANESTHESIA MACHINE
DIVEDED IN TO TWO
      PARTS


• ELECTRICAL SYSTEM

• PNEUMATIC SYSTEM
ELECTRICAL SYSTEM

• Master switch
• Battery back up
• Battery recharge
• Electric outlet for in built
  monitor
• Circuit breaker
PNEUMATIC SYSTEM

• High pressure system

• Intermediate pressure system

• Low pressure system
HIGH PRESSURE

• From cylinders to the pressure
  regulating valve/ preset valve.

• Convert variable pressure flow
  from cylinders to lower and
  most constant pressure
  suitable for machine
• 760 Psi, 1800 Psi to 40-48 Psi
INTERMEDIATE PRESSURE

• Receives gases from pressure
  regulating valve or pipeline &
  supply to the flowmeters &
  oxygen flush valve.

• 40 – 55 Psi from pipeline
• 40 – 48 Psi from pressure
  regulating valve
LOW PRESSURE

• From flow regulating valve to
  common gas outlet.

• Pressure in this system is
  slightly above the atmospheric
  pressure

• 14 -20 Psi
HIGH PRESSURE SYSTEM
HANGER YOKE
 ASSEMBLY
      • Retaining screw

      • Swinging gate
        (Toggle gate
        Swivel gate)

      • Nipple

      • Index pins
7PIN INDEX SYSTEM
       •   machine – pins
       •   Cylinder – holes
       •   Pins – 4 mm in dia
       •   6mm in length, 12°
       •   9/16 inch radius
 7     •   O2 – 2,5
       •   N2O - 3,5
       •   Air – 1,5
       •   Entonox -7
       •   Cyclopropane -3,6
LIMITATIONS OF PIN
        INDEX
• Broken pin

• Upside down yoke block

• Use of double washer
HANGER YOKE
        ASSEMBLY
• WASHER            • FILTER
• GASKET
• Must be present   • Acc to ASTM
• Placed around       standards.
  the nipple
• Only one washer   • 100 µm should
• Two nullify pin     be installed in
  index system        between cylinder
• Must be in good     & reducing valve
  condition
CHECK VALVE ASSEMBLY
• One way valve
• Prevent gas leak from empty
  yoke
• Allow cylinder change while
  using one
• Prevent transfilling of cylinder
• Yoke block, not to keep empty
• cylinder closed – pipeline used.
BOURDEN PRESSURE
     GUAGE




         ELASTIC TUBE
DIFFERENCE IN O2 &
       N2O
PRESSURE
REGULATING VALVE
        • High pressure
          small area, large
          area drop in
          pressure.

        • Spring added for
          constant same
          low pressure
          output.
INTERMIDATE PRESSURE
PIPELINE INLET
 CONNECTION
       • ASTM
       • DISS
       • Check valve
         assembly
       • Pipeline pressure
         indicator gauge
       • 45-55 Psi
       • Cylinder should be
         closed
INTERMIDATE
           PRESSURE
• LEAK –             • GAS POWER
  connection           OUTLET
  inside machine     • Supply to
                       ventilator or jet
• ASTM                 ventilating system
                     • PRESSURE < 50
                       Psi not allowed -
• Leak should not
                       ventilators will not
  be more than 25
                       function
  ml/min at normal
  service pressure   • Quick coupler
                     • DISS
MASTER SWITCH

       • INTERMIDATE
         SYSTEM
       • Downstream the
         inlet
       • Oxygen flush will
         work even if
         standby position
       • Sudden turn off is
         dangerous
OXYGEN PRESSURE
     FAILURE DEVICES
• OXYGN FAILURE        • OXYGEN
  SAFTY – ASTM           FAILURE ALARM
• Set pressure o2      • ASTM
  conc. should not <   • < Set P (30Psi)
  19%                  • Initiated within 5
                         sec .
                       • May be silenced
                         not more than 120
                         sec.
                       • Can’t disabled
OXYGEN PRESSURE
   FAILURE DEVICES
• Drager
• 30-37 Psi 207-255 KPa

• Ohmeda
• 28 Psi 190 KPa
LIMITATIONS
  oxygen failure safety
        device
• System may be pressured for
  weeks if it is gas tight; without
  being connected to the oxygen
  source.

• Do not offer total protection
  against hypoxic mixture
QUICK FLUSH SYSTEM
          • 35 – 75 lit/min
          • 40 - 55 Psi
          • Percutaneous
            catheter –
            adequate
            ventilation
          • Work even when
            master switch off.
          • Collar to prevent
            accidental
            activation.
LIMITATIONS OF
      OXYGEN FLUSH
• ASTM it should     • Flush valve
  not increase or      sticking in ON
  decrease P at        position leads to
  vaporizer outlet
  > 10 KPa           • BAROTRAUMA
  or output > 20 %
                     • AWARENESS
• CHECK VALVE
  IN MACHINE
SECOND STAGE
   REDUCING DEVICE
• Just upstream the N2O & O2
  control valve.
• Gases from pressure reducing
  valve or central supply.
• N2O – 26 Psi
• O2 – 14 Psi
• Flow will be constant below
  normal fluctutation range
FLOW REGULATING
     VALVE
            • Touch, placement,
    PARTS
               color coding.
            • Difference with
              gases.
            • Chemical name
            • Link - mechanical,
              Pneumatic system
            • Pin seat system
            • Stop collar
PROBLEMS WITH FLOW
   CONTROL VALVE
• Inadvertent             • Leak through the
  alteration -              open flow control
   Light                    valve.
  touch, accidental
  brushing                • Failure to allow
                            adequate gas
• Inability to turn the     flow – broken
  control knob ON           stem blocking the
                            path.
LOW PRESSURE SYSTEM
FLOW INDICATER
            • Measurement of
              the flow of the
              gases based on
Lit /min      the principle that
              when gases
              passes through
              resistance there
              is drop in
              pressure
            • Drop in pressure
              corelated with
              flow
FLOWMETER
PHYSICS FLOWMETER

         • LOW FLOWS –
         • LAMINAR FLOW
           VISCOSITY

         • HIGH FLOWS
         • TURBULANT
         • DENSITY
FLOW INDICATER

• RATE OF FLOW          • PRESSURE DROP

• Pressure drop         • Frictional
  across constriction     resistance float &
• Size of annular         tube wall
  opening               • Turbulent flow
• Physical properties   • Drop in energy
  of gas                • Wt float / cross
                          sectional area
FLOW METER
    • Thrope tube
    • Narrow at bottom
    • Calibrated for gas at
      atm. Pressure, 20° C
    • Single, dual taper
    • Antistatic coating
    • Float
    • Float rest, stop
    • Annular space
    • Fluorescent
FLOAT OR BOBBIN
       • Nonrotating
       • Plumb bob, striked
         rotating
       • Ball
       • Indicator of constant
         flow
       • Important to monitor
       • Float stop
       • Float rest
       • Rib guides
PROBLEMS OF
FLOWMETER ASSEMBLY
• FLOWMETER           • FLOAT

• Inaccuracy at low   • Sudden
  flow                  projection to top
• Calibration           with opening of
• Sticking dust         cylinder
• Back pressure       • Block outlet flow
  drops float to      • Wrong flow meter
  lower position.       used.
SEQUENCE OF FLOW
     INDICATER
• Series – one flow control knob
• Such that delivery of hypoxic
  mixture is prevented.
• ASTM – O2 downstream - flow
  indicator – near outlet
• Any separate vaporizer flow
  indicator should be separated
  by at least 10 cm.
IMPORTANCE OF
DOWNSTREAM OF O2
        • Downstream of
          oxygen prevent
           hypoxic mixture
          going to patient in
          case of leak in
          the flow indicator
        • Safety
          mechanism
        • Downstream leak
          change in
          quantity
MINIMUM O2 RATIO
         DEVICES
• Pneumatic systems
• Mechanical systems
• Alarm systems
• Oxygen ratio monitor controller
  & alarm
• Next generation will be
  electronic flow measurement
OHMEDA LINK 25
   SYSTEM
        • Pneumatic -
        • At least 25% of
          oxygen flows in
          the system
        • One of the
          hypoxia guard
        • Mechanical-
          linkage in O2
          N2O knobs by 14
          vs 29 tooth
          sproket
MECHANICAL
O2 RATIO MONITOR
CONTROLLER & ALARM




              Drager
              machine
LOCATION OF
UNIDIRECTIONAL VALVE
UNIDIRECTIONL CHECK
        VALVE
                       • Unidirectional
                         check valve
                       • Between vaporizer
• Positive pressure      & common gas
  from the breathing     outlet
  system is
                       • Upstream of O2
  transmitted to
                         flush flow joins
  machine in IPPV &
                         fresh gas flow
  assisted
  ventilation&O2       • Gas must flow
  flush                  from flowmeter
                       • Leak not detected
PRESSURE RELIEF
        VALVE
• Near the common     • Opens at 100 cm
  gas outlet            of H2O
• To prevent high
  pressure from       • Limits anesthesia
  being transmitted     machine to
  in to machine         deliver jet
• To prevent high       ventilation
  pressure being
  transmitted to Pt   • Present in some
  from machine          machine
VAPORIZER MOUNTING
      DEVICE
• Located in
  between
  flowmeter device
  and common gas
  outlet

• Permanent
  mounting
• Change of
  vaporizer
LOW PRESSURE PIPING

• Large number of connections
• Subject to break or leak

• Flow should not exceed above
  30 ml/min at a pressure of
  3KPa or 30 cm of water
COMMON GAS
         OUTLET
• Receives gases & vapors from
  the machine
• Outlet in most of the machines
  have 15 mm female slip-joint
  connection, with coaxial 22 mm
  male connection
• Frequent location for leak
SAFTY MECHANISMS
• Antistatic coating   • Fluorescent back
• Antistatic wheels      of flow meter
• Color coding         • Master & slaves
• Pin index            • O2 failure alarm
• DISS                 • Oxygen flush
• Pressure guage       • Pressure relief
• Touch, coding, pl      valve
  acement of flow      • Unidirectional
  regulator              check valve
• Rotating bobbin      • Trilene interlock
                         with close system
THANK YOU

Más contenido relacionado

La actualidad más candente (20)

Medical gas supply.pipelines and cylinders
Medical gas supply.pipelines and cylindersMedical gas supply.pipelines and cylinders
Medical gas supply.pipelines and cylinders
 
Anesthesia Machine
Anesthesia MachineAnesthesia Machine
Anesthesia Machine
 
Anaesthesia workstation
Anaesthesia workstationAnaesthesia workstation
Anaesthesia workstation
 
Anaesthesia Vaporizers
Anaesthesia VaporizersAnaesthesia Vaporizers
Anaesthesia Vaporizers
 
Presentation on aneasthetic machine edited 21st july (2)
Presentation on aneasthetic machine edited 21st july (2)Presentation on aneasthetic machine edited 21st july (2)
Presentation on aneasthetic machine edited 21st july (2)
 
Anaesthesia machine 2
Anaesthesia machine 2Anaesthesia machine 2
Anaesthesia machine 2
 
Breathing systems
Breathing systemsBreathing systems
Breathing systems
 
Bougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truviewBougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truview
 
Anesthesia machine
Anesthesia machineAnesthesia machine
Anesthesia machine
 
Anaesthesia machine
Anaesthesia machineAnaesthesia machine
Anaesthesia machine
 
The basic anaesthesia machine
The basic anaesthesia machineThe basic anaesthesia machine
The basic anaesthesia machine
 
Anesthesia Ventilator
Anesthesia VentilatorAnesthesia Ventilator
Anesthesia Ventilator
 
Circle system low flow anesthesia
Circle system low flow anesthesiaCircle system low flow anesthesia
Circle system low flow anesthesia
 
Medical gas cylinders
Medical gas cylindersMedical gas cylinders
Medical gas cylinders
 
Machine boyles
Machine boylesMachine boyles
Machine boyles
 
Retrograde intubation
Retrograde intubationRetrograde intubation
Retrograde intubation
 
#Reservoir bag
#Reservoir bag#Reservoir bag
#Reservoir bag
 
Laryngeal mask-airway
Laryngeal mask-airwayLaryngeal mask-airway
Laryngeal mask-airway
 
Breathing circuits
Breathing circuitsBreathing circuits
Breathing circuits
 
Medical gas Cylinder
Medical gas CylinderMedical gas Cylinder
Medical gas Cylinder
 

Similar a Anesthesia machine

SAFETY FEATURES OF ANAESTHESIA MACHINE
SAFETY FEATURES OF ANAESTHESIA MACHINESAFETY FEATURES OF ANAESTHESIA MACHINE
SAFETY FEATURES OF ANAESTHESIA MACHINEMAHESWARI JAIKUMAR
 
INTERMEDIATE PRESSURE SYSTEM.pptx
INTERMEDIATE PRESSURE SYSTEM.pptxINTERMEDIATE PRESSURE SYSTEM.pptx
INTERMEDIATE PRESSURE SYSTEM.pptxMohitJagga2
 
Anaesthesia machine
Anaesthesia machineAnaesthesia machine
Anaesthesia machinetulsimd
 
Vaporizers types and its working mechanisms
Vaporizers types and its working mechanisms Vaporizers types and its working mechanisms
Vaporizers types and its working mechanisms ZIKRULLAH MALLICK
 
Overview of Conflow Products
Overview of Conflow ProductsOverview of Conflow Products
Overview of Conflow ProductsGary Wain
 
PressureRegulators.ppt
PressureRegulators.pptPressureRegulators.ppt
PressureRegulators.pptaruncs28
 
Intermediate & low pressure system
Intermediate & low pressure systemIntermediate & low pressure system
Intermediate & low pressure systemDeepa Sinha
 
anaesthesia machine-140810030802-pptx.pptx
anaesthesia machine-140810030802-pptx.pptxanaesthesia machine-140810030802-pptx.pptx
anaesthesia machine-140810030802-pptx.pptxRaj Kumar
 
flow measurement.ppt
flow measurement.pptflow measurement.ppt
flow measurement.pptsibtikhokhar
 
Control Valve Types
Control Valve TypesControl Valve Types
Control Valve TypesBrannon Gant
 
Pre commissioning steam turbines load trial
Pre commissioning steam turbines load trialPre commissioning steam turbines load trial
Pre commissioning steam turbines load trialNagesh H
 
Oxygen MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECT
Oxygen  MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECTOxygen  MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECT
Oxygen MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECTDr.RMLIMS lucknow
 
Ball & tube mill
Ball & tube millBall & tube mill
Ball & tube millStudy Hub
 
Oil & Gas_Process Control Market FCDE 2015
Oil & Gas_Process Control Market FCDE 2015Oil & Gas_Process Control Market FCDE 2015
Oil & Gas_Process Control Market FCDE 2015Blessing Chioma Udongwo
 
CONTROL VALVE questions.pdf
CONTROL VALVE questions.pdfCONTROL VALVE questions.pdf
CONTROL VALVE questions.pdfAmiTG34
 

Similar a Anesthesia machine (20)

SAFETY FEATURES OF ANAESTHESIA MACHINE
SAFETY FEATURES OF ANAESTHESIA MACHINESAFETY FEATURES OF ANAESTHESIA MACHINE
SAFETY FEATURES OF ANAESTHESIA MACHINE
 
INTERMEDIATE PRESSURE SYSTEM.pptx
INTERMEDIATE PRESSURE SYSTEM.pptxINTERMEDIATE PRESSURE SYSTEM.pptx
INTERMEDIATE PRESSURE SYSTEM.pptx
 
Control valve
Control valveControl valve
Control valve
 
Anesthesia Machine -.pptx
Anesthesia Machine -.pptxAnesthesia Machine -.pptx
Anesthesia Machine -.pptx
 
Anaesthesia machine
Anaesthesia machineAnaesthesia machine
Anaesthesia machine
 
Vaporizers types and its working mechanisms
Vaporizers types and its working mechanisms Vaporizers types and its working mechanisms
Vaporizers types and its working mechanisms
 
Overview of Conflow Products
Overview of Conflow ProductsOverview of Conflow Products
Overview of Conflow Products
 
PressureRegulators.ppt
PressureRegulators.pptPressureRegulators.ppt
PressureRegulators.ppt
 
Intermediate & low pressure system
Intermediate & low pressure systemIntermediate & low pressure system
Intermediate & low pressure system
 
anaesthesia machine-140810030802-pptx.pptx
anaesthesia machine-140810030802-pptx.pptxanaesthesia machine-140810030802-pptx.pptx
anaesthesia machine-140810030802-pptx.pptx
 
flow measurement.ppt
flow measurement.pptflow measurement.ppt
flow measurement.ppt
 
Basic instrumentation
Basic instrumentationBasic instrumentation
Basic instrumentation
 
Basic instrumentation
Basic instrumentationBasic instrumentation
Basic instrumentation
 
Actuator and valves
Actuator and valvesActuator and valves
Actuator and valves
 
Control Valve Types
Control Valve TypesControl Valve Types
Control Valve Types
 
Pre commissioning steam turbines load trial
Pre commissioning steam turbines load trialPre commissioning steam turbines load trial
Pre commissioning steam turbines load trial
 
Oxygen MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECT
Oxygen  MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECTOxygen  MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECT
Oxygen MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECT
 
Ball & tube mill
Ball & tube millBall & tube mill
Ball & tube mill
 
Oil & Gas_Process Control Market FCDE 2015
Oil & Gas_Process Control Market FCDE 2015Oil & Gas_Process Control Market FCDE 2015
Oil & Gas_Process Control Market FCDE 2015
 
CONTROL VALVE questions.pdf
CONTROL VALVE questions.pdfCONTROL VALVE questions.pdf
CONTROL VALVE questions.pdf
 

Anesthesia machine

  • 1. ANESTHESIA MACHINE DR. PRIYANKA
  • 2. ANESTHESIA MACHINE • The function of anesthesia machine is preparation of anesthesia gas mixture of pure and precisely known yet variable composition at controlled & known pressure.
  • 3. MACHINE STANDARDS • 1979 – ANSI • 1988 – ASTM • Defined design, performance, safety requirements of anesthesia machine.
  • 6. DIVEDED IN TO TWO PARTS • ELECTRICAL SYSTEM • PNEUMATIC SYSTEM
  • 7. ELECTRICAL SYSTEM • Master switch • Battery back up • Battery recharge • Electric outlet for in built monitor • Circuit breaker
  • 8. PNEUMATIC SYSTEM • High pressure system • Intermediate pressure system • Low pressure system
  • 9. HIGH PRESSURE • From cylinders to the pressure regulating valve/ preset valve. • Convert variable pressure flow from cylinders to lower and most constant pressure suitable for machine • 760 Psi, 1800 Psi to 40-48 Psi
  • 10. INTERMEDIATE PRESSURE • Receives gases from pressure regulating valve or pipeline & supply to the flowmeters & oxygen flush valve. • 40 – 55 Psi from pipeline • 40 – 48 Psi from pressure regulating valve
  • 11. LOW PRESSURE • From flow regulating valve to common gas outlet. • Pressure in this system is slightly above the atmospheric pressure • 14 -20 Psi
  • 13. HANGER YOKE ASSEMBLY • Retaining screw • Swinging gate (Toggle gate Swivel gate) • Nipple • Index pins
  • 14. 7PIN INDEX SYSTEM • machine – pins • Cylinder – holes • Pins – 4 mm in dia • 6mm in length, 12° • 9/16 inch radius 7 • O2 – 2,5 • N2O - 3,5 • Air – 1,5 • Entonox -7 • Cyclopropane -3,6
  • 15. LIMITATIONS OF PIN INDEX • Broken pin • Upside down yoke block • Use of double washer
  • 16. HANGER YOKE ASSEMBLY • WASHER • FILTER • GASKET • Must be present • Acc to ASTM • Placed around standards. the nipple • Only one washer • 100 µm should • Two nullify pin be installed in index system between cylinder • Must be in good & reducing valve condition
  • 17. CHECK VALVE ASSEMBLY • One way valve • Prevent gas leak from empty yoke • Allow cylinder change while using one • Prevent transfilling of cylinder • Yoke block, not to keep empty • cylinder closed – pipeline used.
  • 18. BOURDEN PRESSURE GUAGE ELASTIC TUBE
  • 20. PRESSURE REGULATING VALVE • High pressure small area, large area drop in pressure. • Spring added for constant same low pressure output.
  • 22. PIPELINE INLET CONNECTION • ASTM • DISS • Check valve assembly • Pipeline pressure indicator gauge • 45-55 Psi • Cylinder should be closed
  • 23. INTERMIDATE PRESSURE • LEAK – • GAS POWER connection OUTLET inside machine • Supply to ventilator or jet • ASTM ventilating system • PRESSURE < 50 Psi not allowed - • Leak should not ventilators will not be more than 25 function ml/min at normal service pressure • Quick coupler • DISS
  • 24. MASTER SWITCH • INTERMIDATE SYSTEM • Downstream the inlet • Oxygen flush will work even if standby position • Sudden turn off is dangerous
  • 25. OXYGEN PRESSURE FAILURE DEVICES • OXYGN FAILURE • OXYGEN SAFTY – ASTM FAILURE ALARM • Set pressure o2 • ASTM conc. should not < • < Set P (30Psi) 19% • Initiated within 5 sec . • May be silenced not more than 120 sec. • Can’t disabled
  • 26. OXYGEN PRESSURE FAILURE DEVICES • Drager • 30-37 Psi 207-255 KPa • Ohmeda • 28 Psi 190 KPa
  • 27. LIMITATIONS oxygen failure safety device • System may be pressured for weeks if it is gas tight; without being connected to the oxygen source. • Do not offer total protection against hypoxic mixture
  • 28. QUICK FLUSH SYSTEM • 35 – 75 lit/min • 40 - 55 Psi • Percutaneous catheter – adequate ventilation • Work even when master switch off. • Collar to prevent accidental activation.
  • 29. LIMITATIONS OF OXYGEN FLUSH • ASTM it should • Flush valve not increase or sticking in ON decrease P at position leads to vaporizer outlet > 10 KPa • BAROTRAUMA or output > 20 % • AWARENESS • CHECK VALVE IN MACHINE
  • 30. SECOND STAGE REDUCING DEVICE • Just upstream the N2O & O2 control valve. • Gases from pressure reducing valve or central supply. • N2O – 26 Psi • O2 – 14 Psi • Flow will be constant below normal fluctutation range
  • 31. FLOW REGULATING VALVE • Touch, placement, PARTS color coding. • Difference with gases. • Chemical name • Link - mechanical, Pneumatic system • Pin seat system • Stop collar
  • 32. PROBLEMS WITH FLOW CONTROL VALVE • Inadvertent • Leak through the alteration - open flow control Light valve. touch, accidental brushing • Failure to allow adequate gas • Inability to turn the flow – broken control knob ON stem blocking the path.
  • 34. FLOW INDICATER • Measurement of the flow of the gases based on Lit /min the principle that when gases passes through resistance there is drop in pressure • Drop in pressure corelated with flow
  • 36. PHYSICS FLOWMETER • LOW FLOWS – • LAMINAR FLOW VISCOSITY • HIGH FLOWS • TURBULANT • DENSITY
  • 37. FLOW INDICATER • RATE OF FLOW • PRESSURE DROP • Pressure drop • Frictional across constriction resistance float & • Size of annular tube wall opening • Turbulent flow • Physical properties • Drop in energy of gas • Wt float / cross sectional area
  • 38. FLOW METER • Thrope tube • Narrow at bottom • Calibrated for gas at atm. Pressure, 20° C • Single, dual taper • Antistatic coating • Float • Float rest, stop • Annular space • Fluorescent
  • 39. FLOAT OR BOBBIN • Nonrotating • Plumb bob, striked rotating • Ball • Indicator of constant flow • Important to monitor • Float stop • Float rest • Rib guides
  • 40. PROBLEMS OF FLOWMETER ASSEMBLY • FLOWMETER • FLOAT • Inaccuracy at low • Sudden flow projection to top • Calibration with opening of • Sticking dust cylinder • Back pressure • Block outlet flow drops float to • Wrong flow meter lower position. used.
  • 41. SEQUENCE OF FLOW INDICATER • Series – one flow control knob • Such that delivery of hypoxic mixture is prevented. • ASTM – O2 downstream - flow indicator – near outlet • Any separate vaporizer flow indicator should be separated by at least 10 cm.
  • 42. IMPORTANCE OF DOWNSTREAM OF O2 • Downstream of oxygen prevent hypoxic mixture going to patient in case of leak in the flow indicator • Safety mechanism • Downstream leak change in quantity
  • 43. MINIMUM O2 RATIO DEVICES • Pneumatic systems • Mechanical systems • Alarm systems • Oxygen ratio monitor controller & alarm • Next generation will be electronic flow measurement
  • 44. OHMEDA LINK 25 SYSTEM • Pneumatic - • At least 25% of oxygen flows in the system • One of the hypoxia guard • Mechanical- linkage in O2 N2O knobs by 14 vs 29 tooth sproket
  • 46. O2 RATIO MONITOR CONTROLLER & ALARM Drager machine
  • 48. UNIDIRECTIONL CHECK VALVE • Unidirectional check valve • Between vaporizer • Positive pressure & common gas from the breathing outlet system is • Upstream of O2 transmitted to flush flow joins machine in IPPV & fresh gas flow assisted ventilation&O2 • Gas must flow flush from flowmeter • Leak not detected
  • 49. PRESSURE RELIEF VALVE • Near the common • Opens at 100 cm gas outlet of H2O • To prevent high pressure from • Limits anesthesia being transmitted machine to in to machine deliver jet • To prevent high ventilation pressure being transmitted to Pt • Present in some from machine machine
  • 50. VAPORIZER MOUNTING DEVICE • Located in between flowmeter device and common gas outlet • Permanent mounting • Change of vaporizer
  • 51. LOW PRESSURE PIPING • Large number of connections • Subject to break or leak • Flow should not exceed above 30 ml/min at a pressure of 3KPa or 30 cm of water
  • 52. COMMON GAS OUTLET • Receives gases & vapors from the machine • Outlet in most of the machines have 15 mm female slip-joint connection, with coaxial 22 mm male connection • Frequent location for leak
  • 53. SAFTY MECHANISMS • Antistatic coating • Fluorescent back • Antistatic wheels of flow meter • Color coding • Master & slaves • Pin index • O2 failure alarm • DISS • Oxygen flush • Pressure guage • Pressure relief • Touch, coding, pl valve acement of flow • Unidirectional regulator check valve • Rotating bobbin • Trilene interlock with close system