ANUBHAV GUPTA
B. PHARM 4TH YEAR
Institute of Pharmacy
VBS Purvanchal University
1.INTRODUCTION
▪ Mass Spectrometry (MS) is a technique that helps to identify the amount and type of chemicals
present in a sample by measuring the mass-to-charge ratio and abundance of gas-phase ions.
▪ A mass spectrometer works by generating charged molecules or molecular fragments either in a
high vacuum or immediately prior to the sample entering the high-vacuum region.
▪ Mass spectrum is a plot of relative abundance against the ratio of mass/charge (m/e).
▪ These spectra are used to determine the elemental or isotopic signature of a sample, the masses of
particles and of molecules, and to elucidate the chemical structures of molecules and other chemical
compounds.
▪ Mass spectrometry provides a highly specific method for determining or confirming the identity or
structure of drugs and raw materials used in their manufacture.
▪ Mass spectrometry in conjunction with either gas chromatography (GC–MS) or liquid
chromatography (LC–MS) provides a method for characterising impurities in drugs and formulation
excipients.
2
Anubhav Gupta
2.PRINCIPLE
▪ The mass spectroscopy works on principle that when charged molecules or molecular fragments are
generated in a high-vacuum region, or immediately prior to a sample entering a high-vacuum
region, using a variety of methods for ion production.
▪ The ions are generated in the gas phase so that they can then be manipulated by the application of
either electric or magnetic fields to enable the determination of their molecular weights.
3
Anubhav Gupta
3.Working of Mass Spectrometry
▪ In a typical procedure, a sample, which may be solid, liquid, or gas, is ionized, for example by
bombarding it with electrons.
▪ This may cause some of the sample’s molecules to break into charged fragments. These ions are
then separated according to their mass-to-charge ratio, typically by accelerating them and
subjecting them to an electric or magnetic field:
▪ Ions of the same mass-to-charge ratio will undergo the same amount of deflection.
▪ The ions are detected by a mechanism capable of detecting charged particles, such as an electron
multiplier.
▪ Results are displayed as spectra of the relative abundance of detected ions as a function of the
mass-to-charge ratio.
▪ The atoms or molecules in the sample can be identified by correlating known masses (e.g. an
entire molecule) to the identified masses or through a characteristic fragmentation pattern.
4
Anubhav Gupta
4.FRAGMENTATION PATTERN
▪ Fragmentation is the dissociation of energetically unstable molecular ions formed from passing
the molecules in the ionization chamber of a mass spectrometer.
▪ Fragmentation of the molecular ions takes place in following forms:
1. Homolytic cleavage. This type of fragmentation is promoted by the presence of a hetero atom
such as oxygen, nitrogen or sulphur, and in molecules containing a hetero atom it often gives rise
to the most abundant ion in the mass spectrum
2. Heterolytic cleavage. Heterolytic cleavage occurs less often as a predominant fragmentation
mechanism in drug molecules since they usually contain a lot of heteroatoms to direct the
cleavage. In such cleavage, the positive charge is carried by the carbon atom not by the
heteroatom.
3. Retro Diels-Alders reaction. It occurs in compounds with ring systems. It involves the
cleavage of two bonds of a cyclic system which results in the formation of two stable
unsaturated fragment in which two new bonds are formed.
5
Anubhav Gupta
4.1.Fragmentation Pattern
ALOCOHOL(3-Pentanol) ALDEHYDE(3-Phenyl 2-Propenal)
ALKANE(Hexane) AMIDE(Methyl butyramide)
6
Anubhav Gupta
5.McLafferty Rearrangement
▪ The McLafferty rearrangement is a characteristic fragmentation of the molecular ion of a carbonyl
compound containing at least one gamma hydrogen, e.g., Migration of -hydrogen followed by
-bond cleavage and elimination of ethylene or substituted ethylene neutral molecule.
▪ The McLafferty rearrangement, is relatively uncommon
in drug molecules; it is most often encountered in fatty
acid esters.
▪ This rearrangement can occur in carbonyl compounds
e.g. amides, ketones and acids.
7
Anubhav Gupta
6.IONIZATION TECHNIQUES
▪ There are many types of ionization techniques that are used in mass spectrometry. Some of them
are:
1.Electron Impact ionization (EI)
2.Fast Atom Bombardment (FAB)
3.Electrospray ionization (ESI)
4.Atmospheric Pressure Chemical Ionization (APCI)
5.Matrix Assisted Laser Desorption Ionization (MALDI)
▪ APCI is considered best than EI because APCI forms a protonated molecule and is completely
compatible with Liquid Chromatography (LC), while EI is more likely to fragment the ion leading
to a possible more ambiguous identification of the molecule weight and is incompatible with LC.
▪ MALDI along with ESI allows for ionization and measurement of large molecular weight.
▪ ESI has an advantage in its easy compatibility with LC. While MALDI has advantages for imaging
mass spectrometry.
8
Anubhav Gupta
9
6.1. Electron Impact Ionization(EI)
▪ EI is still used in conjunction with sample introduction either via a direct heated probe or via gas
chromatography (GC):
(i) The sample is introduced into the instrument source by heating it on the end of a probe until it
evaporates, assisted by the high vacuum within the instrument or via a capillary GC column.
(ii) Once in the vapour phase, the analyte is bombarded with the electrons produced by a rhenium or
tungsten filament, which are accelerated towards a positive target with an energy of 70 eV. The analyte is
introduced between the filament and the target, and the electrons cause
ionisation as follows: M þ e ! Mþ: þ 2 e.
(iii) Since the electrons used are of much higher energy than the
strength of the bonds within the analyte (4–7 eV), extensive
fragmentation of the analyte usually occurs.
(iv) The molecule and its fragments are pushed out of the source
by a repeller plate which has the same charge as the ions
generated.
Anubhav Gupta
10
6.2. Fast Atom Bombardment(FAB)
▪ FAB is a technique that was popular in the 80's to early 90's because it was the first technique that
allowed ionization of non-volatile compounds that could be done simply.
▪ It was done by bombarding a sample in a vacuum with a beam of atoms, typically Ar or Xe,
accelerated to Kilovolt energies. The sample was typically mixed in a matrix.
▪ The two most common matrixes were glycerol and 3 Nitro-benzoic acid. The matrix allowed the
sample to refresh itself.
▪ The ions formed by FAB were adducts to the molecule, where the
adducts could be protons, sodium ions, potassium ions or
ammonium ions.
▪ A variation of FAB was replacement of the atom beam with a
beam of ions, typically cesium ions, which was called secondary
ion mass spectrometry (SIMS). SIMS spectra were typically
identical to FAB spectra and the terms became interchangeable.
Anubhav Gupta
11
6.3. Electronspray Ionization(ESI)
▪ The electronspray is created by putting a high voltage on a flow of liquid at atmospheric pressure,
sometimes this is assisted by a concurrent flow of gas.
▪ The created spray is directed to an opening in the vacuum system of the mass spectrometer, where
the droplets are de-solvated by a combination of heat, vacuum and acceleration into gas by voltages.
▪ The ions are ejected in droplets and accelerated into
the mass analyzer by voltages. For larger molecules,
the ions may contain multiple charges, allowing the
detection of very large molecules on analyzers that
have limited mass to charge (m/Z)) ratio ranges.
Because of the natural use of a flowing liquid, it is
easily adapted to liquid chromatography (LC).
▪ There are many other techniques that are variations of
electrospray, for example, nanospray, picospray
desorption electrospray ionization (DESI).
Anubhav Gupta
12
6.4. Atmospheric Pressure Chemical Ionization (APCI)
▪ Atmospheric pressure chemical ionisation (APCI) is closely related to ESI and instruments carrying
out ESI can be readily switched to operate in APCI mode.
▪ In APCI mode the eluent from the HPLC does not pass through a charged needle before entering
the mass spectrometer source but via a heated tube so that it forms an aerosol. Upon exiting the
heated tube an electric discharge is passed through the aerosol generating reactive species such as
H3Oþ and N2þ, which promote the ionisation of the
analytes.
▪ The technique has never achieved the level of
popularity of ESI since most drug molecules can be
ionised under ESI conditions; however, APCI can be
employed for the analysis of drug molecules of low
polarity that do not ionise efficiently under ESI
conditions.
Anubhav Gupta
13
6.5. Matrix Assisted Laser Desorption Ionization (MALDI)
▪ MALDI uses a nitrogen laser to promote ionisation of molecules prior to ion separation in a mass
spectrometer. It is usually combined with time of flight (TOF) separation of the ions generated.
▪ In order for the sample to be ionised it has to be dissolved in a matrix that absorbs UV radiation at
around the wavelength (337 nm) produced by the laser. A simple example of a matrix is
dihydroxybenzoic acid and there are a number of similar aromatic compounds which are used to
promote ionisation of different classes of molecules.
▪ The sample solution is mixed with matrix solution on a metal plate and allowed to dry prior to being
introduced into the instrument. The laser is then directed at the target plate to promote ionisation.
Anubhav Gupta
14
7. INSTRUMENTATION
Mass spectrometer has following components:
A. Sample Inlet, through HPLC, GC, Syringe etc.
B. Ionization, can be achieved by :
1.Electron Impact ionization (EI)
2.Fast Atom Bombardment (FAB)
3.Electrospray ionization (ESI)
4.Atmospheric Pressure Chemical Ionization (APCI)
5.Matrix Assisted Laser Desorption Ionization (MALDI)
C. Acceleration & Deflection
D. Analyser, can be achieved by analysers like :
1.Magnetic sector mass analyser
2.Double focussing analyser
3.Quadrupole mass analyser
4.Time of Flight analyser (TOF)
5.Ion trap analyser
6.Ion cyclotron analyser
E. Detector
Anubhav Gupta
15
7.1. Sample Inlet
▪ Sample stored in large reservoir from which molecules reaches ionization chamber at low
pressure in steady stream by a pinhole called “Molecular leak”.
▪ Molecular leak–It is pin-hole restriction (0.01 to 0.05mm diameter) and made up of gold foil. It
is used for metering the sample to ionization chamber.
7.2. Ionization
▪ Atoms are ionized by knocking one or more electrons off to give positive ions by bombardment
with a stream of electrons. Most of the positive ions formed will carry charge of +1.
7.3.Acceleration & Deflection
▪ Ions are accelerated so that they all have same kinetic energy.
▪ Positive ions pass through 3 slits with voltage in decreasing order.
▪ Middle slit carries intermediate and finals at zero volts.
▪ Ions are deflected by a magnetic field due to difference in their masses.
▪ The lighter the mass, more they are deflected.
▪ It also depends upon the no. of +ve charge an ion is carrying; the more +ve charge, more it will be
deflected.
Anubhav Gupta
16
7.4. Analyser
▪ A mass analyser is a device that can separate atoms and molecules according to their mass.
▪ The five main characteristics for measuring the performance of a mass analyser are
1) The mass range limit or dynamic range
2) The analysis speed [u (m)S-1]
3) The transmission = No. of ion reaching the ions/No. of ions entering mass analyzer
4) The mass accuracy
5) The resolution.
▪ Mass analysers used are:
1.Magnetic sector mass analyser
2.Double focussing analyser
3.Quadrupole mass analyser
4.Time of Flight analyser (TOF)
5.Ion trap analyser
6.Ion cyclotron analyser
Anubhav Gupta
17
7.4.1. Quadrupole Mass Analyzer
▪ A cheaper and more sensitive mass spectrometer than a magnetic sector instrument is based on the
quadrupole analyser.
▪ It uses two electric fields applied at right angles to each other, rather than a magnetic field, to
separate ions according to their m/z ratios. One of the fields used is DC and the other oscillates at
radiofrequency.
Disadvantage
▪ Limited resolution
▪ Peak heights variable as a function of mass (mass
discrimination).
▪ Peak height vs. mass response must be 'tuned'.
▪ Not well suited for pulsed ionization methods.
▪ Low-energy collision-induced dissociation (CID)
MS/MS spectra in triple quadrupole and hybrid
mass spectrometers depend strongly on energy,
collision gas, pressure, and other factors.
Anubhav Gupta
18
7.4.2. Time of flight (TOF) analyser
▪ The basis of the separation is that smaller ions move more quickly than larger ions and thus reach the
detector first.
▪ The technique relies on gating the signal from the ion source, so that one packet of ions is allowed
time to reach the detector before the next set is ejected from the ion source, so that there is no
overlap between ion packets. The ions leaving the ion source have different kinetic energies and this
compromises the mass resolution, indeed early instruments gave very wide peaks for a single mass.
▪ The problem of mass resolution was resolved by
using a device called a reflectron, which opposes
the direction that the ions are moving by sending
them back in the opposite direction.
▪ The faster moving ions penetrate deeper into the
reflectron and thus a lag effect is produced for
faster moving ions so that they are focused with
the slower moving ions which do not penetrate as
far into the reflectron. Pushing the ions back in the
opposite direction also increases the length of the
flight tube, increasing instrument resolution.
Anubhav Gupta
19
7.4.3. Magnetic Sector Mass Analyzer
▪ In a magnetic sector instrument the ions generated are pushed out of the source by a repeller
potential of same charge as the ion itself (most often positive).
▪ They are then accelerated in an electric field of ca 3–8 kV and travel through an electrostatic field
region so that they are forced to fall into a narrow range of kinetic energies prior to entering the
field of a circular magnet.
▪ They then adopt a flight path through the magnetic field depending on their charge to mass (m/z)
ratio; the large ions are deflected less by the magnetic field:
𝒎
𝒛
=
𝑯𝟐𝒓𝟐
𝟐𝑽
where H is the magnetic field strength, r is the
radius of the circular path in which the ion travels,
and V is the accelerating voltage.
Anubhav Gupta
20
Comparison of various analyzers
Anubhav Gupta
21
7.5. Detectors
▪ The ion collection system measures the relative abundance of ion fragments of each mass.
▪ Several types of detectors are available for mass spectrometers. The detector used for most routine
experiments is the electron multiplier.
▪ Another type of detector is photographic plates coated with a silver bromide emulsion, it is sensitive
to energetic ions. A photographic plate can give a higher resolution than an electrical detector.
Faraday Cup:
▪ It consists of a hollow conducting electrode
connected to ground via a high resistance.
▪ The ions hitting the collector cause a flow of
electrons from ground to the resistor.
▪ The resulting potential drop across the resistor
is amplified. A plate held at about -80 V in
front of the collector, prevents any ejected
secondary electrons from escaping and
causing an anomalous reading.
Photographic Plates:
▪ This detector system is most sensitive than any
other detector because the photoplate
integrates the ion signal over a period of time.
▪ The photoplate are processed by the usual
photographic techniques and read with the aid
of densitometer.
Anubhav Gupta
22
Electron Multipliers:
▪ It is most common means of detecting ions. It is made up of a series (12 to 24) of aluminum oxide
dynodes maintained at ever increasing potentials.
▪ Ions strike the first dynode surface causing an emission of electrons. These electrons are then
attracted to the next dynode held at a higher potential and therefore more secondary electrons are
generated.
▪ Ultimately, as numerous dynodes are involved, a cascade of electrons is formed that reult in an
overall current gain on the order of one million or higher.
▪ The high energy dynode(HED) uses an electrostatic field to increase the velocity of the ions and
serves to increase signal intensity and therefore sensitivity.
Faraday Cup Photoplate Electron Multiplier
Anubhav Gupta
23
8.APPLICATIONS
▪ Mass spectrometry provides a highly specific method for determining or confirming the identity or
structure of drugs and raw materials used in their manufacture.
▪ Mass spectrometry in conjunction with either gas chromatography (GC–MS) or liquid
chromatography (LC–MS) provides a method for characterising impurities in drugs and
formulation excipients.
▪ Pharmaceutical Analysis
Bioavailability studies
Drug metabolism studies
Characterization of potential drugs
Drug degradation product analysis
Identifying drug targets
▪ Biomolecular characterization
Characterization of Proteins an Peptides
Oligonucleotides.
▪ Environmental Analysis
▪ Forensic Analysis
Anubhav Gupta

Más contenido relacionado

La actualidad más candente

Molecular spectroscopy 1
Molecular spectroscopy 1Molecular spectroscopy 1
Molecular spectroscopy 1Dewal Deshmukh
 
Mass spectroscopy and fragmentation rule
Mass spectroscopy and fragmentation rule Mass spectroscopy and fragmentation rule
Mass spectroscopy and fragmentation rule Zahid Wajdan
 
factors affecting fluorescence & phosphorescence
 factors affecting fluorescence & phosphorescence factors affecting fluorescence & phosphorescence
factors affecting fluorescence & phosphorescenceRamsha Afzal
 
instrumentation of mass spectrometry
instrumentation of mass spectrometryinstrumentation of mass spectrometry
instrumentation of mass spectrometryManali Parab
 
UV-visible spectroscopy
UV-visible spectroscopyUV-visible spectroscopy
UV-visible spectroscopyaaaaalok
 
Hyphenated techniques
Hyphenated techniquesHyphenated techniques
Hyphenated techniquesBiji Saro
 
Mass Spectroscopy and its applications
Mass Spectroscopy and its applicationsMass Spectroscopy and its applications
Mass Spectroscopy and its applicationsAchyut Adhikari
 
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPYINDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPYParimi Anuradha
 
MASS SPECTROSCOPY & ITS INSTRUMENTATION
MASS SPECTROSCOPY & ITS INSTRUMENTATIONMASS SPECTROSCOPY & ITS INSTRUMENTATION
MASS SPECTROSCOPY & ITS INSTRUMENTATIONROHIT
 
Chemical ionization
Chemical ionizationChemical ionization
Chemical ionizationAarif Khan
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometryarushe143
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopyUlivin Al Farisi
 
Ultraviolet (uv) and visible spectroscopy ppt
Ultraviolet (uv) and visible spectroscopy pptUltraviolet (uv) and visible spectroscopy ppt
Ultraviolet (uv) and visible spectroscopy pptBharathiar university
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopyRAHEELA Khan
 

La actualidad más candente (20)

Molecular spectroscopy 1
Molecular spectroscopy 1Molecular spectroscopy 1
Molecular spectroscopy 1
 
Mass spectroscopy and fragmentation rule
Mass spectroscopy and fragmentation rule Mass spectroscopy and fragmentation rule
Mass spectroscopy and fragmentation rule
 
factors affecting fluorescence & phosphorescence
 factors affecting fluorescence & phosphorescence factors affecting fluorescence & phosphorescence
factors affecting fluorescence & phosphorescence
 
Hyphenated techniques
Hyphenated techniquesHyphenated techniques
Hyphenated techniques
 
instrumentation of mass spectrometry
instrumentation of mass spectrometryinstrumentation of mass spectrometry
instrumentation of mass spectrometry
 
Raman spectroscopy
Raman spectroscopyRaman spectroscopy
Raman spectroscopy
 
UV-visible spectroscopy
UV-visible spectroscopyUV-visible spectroscopy
UV-visible spectroscopy
 
LC- MS
LC- MS LC- MS
LC- MS
 
Hyphenated techniques
Hyphenated techniquesHyphenated techniques
Hyphenated techniques
 
Mass Spectroscopy and its applications
Mass Spectroscopy and its applicationsMass Spectroscopy and its applications
Mass Spectroscopy and its applications
 
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPYINDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
 
Electron spectroscopy
Electron spectroscopyElectron spectroscopy
Electron spectroscopy
 
MASS SPECTROSCOPY & ITS INSTRUMENTATION
MASS SPECTROSCOPY & ITS INSTRUMENTATIONMASS SPECTROSCOPY & ITS INSTRUMENTATION
MASS SPECTROSCOPY & ITS INSTRUMENTATION
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometry
 
Chemical ionization
Chemical ionizationChemical ionization
Chemical ionization
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometry
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
MASS SPECTROMETRY=.pdf
MASS SPECTROMETRY=.pdfMASS SPECTROMETRY=.pdf
MASS SPECTROMETRY=.pdf
 
Ultraviolet (uv) and visible spectroscopy ppt
Ultraviolet (uv) and visible spectroscopy pptUltraviolet (uv) and visible spectroscopy ppt
Ultraviolet (uv) and visible spectroscopy ppt
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 

Similar a mass spectroscopy

Mass.pptx instrumentation, principle, theory
Mass.pptx instrumentation, principle, theoryMass.pptx instrumentation, principle, theory
Mass.pptx instrumentation, principle, theoryDr. Vijaya Barge
 
Ionization Techniques in Mass Spectrometry.pdf
Ionization Techniques in Mass Spectrometry.pdfIonization Techniques in Mass Spectrometry.pdf
Ionization Techniques in Mass Spectrometry.pdfAli Hamza
 
Mass Ionization Techniques
Mass Ionization TechniquesMass Ionization Techniques
Mass Ionization TechniquesMehulJain143
 
Mass Spectrometry Ionization Techniques
Mass Spectrometry Ionization TechniquesMass Spectrometry Ionization Techniques
Mass Spectrometry Ionization TechniquesReyaz007
 
mass spectrometery
mass spectrometerymass spectrometery
mass spectrometeryuog
 
Lc -ms seminar
Lc -ms seminarLc -ms seminar
Lc -ms seminarNIDHIYJ
 
Protein mass spectrometry data analysis.
Protein mass spectrometry data analysis.Protein mass spectrometry data analysis.
Protein mass spectrometry data analysis.NahidRehman
 
atomic absorption spectroscopy and mass spectroscopy
atomic absorption spectroscopy and mass spectroscopyatomic absorption spectroscopy and mass spectroscopy
atomic absorption spectroscopy and mass spectroscopyvanitha gopal
 
Information about the ionization spectroscopy technique
Information about the ionization spectroscopy techniqueInformation about the ionization spectroscopy technique
Information about the ionization spectroscopy techniquetehreemusman2021
 
Liquid chromatography–mass spectrometry (LC-MS) BY P. RAVISANKAR
Liquid chromatography–mass spectrometry (LC-MS) BY P. RAVISANKARLiquid chromatography–mass spectrometry (LC-MS) BY P. RAVISANKAR
Liquid chromatography–mass spectrometry (LC-MS) BY P. RAVISANKARDr. Ravi Sankar
 
Bilal Ibrahim Dan-Iya_GS57595.pptx
Bilal Ibrahim Dan-Iya_GS57595.pptxBilal Ibrahim Dan-Iya_GS57595.pptx
Bilal Ibrahim Dan-Iya_GS57595.pptxDan-Iya
 
liquid chromatography-mass spectrometry
liquid chromatography-mass spectrometryliquid chromatography-mass spectrometry
liquid chromatography-mass spectrometrysamiya shaik
 

Similar a mass spectroscopy (20)

Prabhakar singh ii sem-paper v-mass spectroscopy
Prabhakar singh  ii sem-paper v-mass spectroscopyPrabhakar singh  ii sem-paper v-mass spectroscopy
Prabhakar singh ii sem-paper v-mass spectroscopy
 
Mass.pptx instrumentation, principle, theory
Mass.pptx instrumentation, principle, theoryMass.pptx instrumentation, principle, theory
Mass.pptx instrumentation, principle, theory
 
Mass assingment
Mass  assingmentMass  assingment
Mass assingment
 
Mass spectrometry i
Mass spectrometry iMass spectrometry i
Mass spectrometry i
 
Mass spectroscopy pdf
Mass spectroscopy  pdfMass spectroscopy  pdf
Mass spectroscopy pdf
 
Ionization Techniques in Mass Spectrometry.pdf
Ionization Techniques in Mass Spectrometry.pdfIonization Techniques in Mass Spectrometry.pdf
Ionization Techniques in Mass Spectrometry.pdf
 
Mass Spectrometry
Mass SpectrometryMass Spectrometry
Mass Spectrometry
 
Mass Ionization Techniques
Mass Ionization TechniquesMass Ionization Techniques
Mass Ionization Techniques
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometry
 
Mass Spectrometry Ionization Techniques
Mass Spectrometry Ionization TechniquesMass Spectrometry Ionization Techniques
Mass Spectrometry Ionization Techniques
 
mass spectrometery
mass spectrometerymass spectrometery
mass spectrometery
 
59 330-l2-ms1
59 330-l2-ms159 330-l2-ms1
59 330-l2-ms1
 
Lc -ms seminar
Lc -ms seminarLc -ms seminar
Lc -ms seminar
 
Protein mass spectrometry data analysis.
Protein mass spectrometry data analysis.Protein mass spectrometry data analysis.
Protein mass spectrometry data analysis.
 
atomic absorption spectroscopy and mass spectroscopy
atomic absorption spectroscopy and mass spectroscopyatomic absorption spectroscopy and mass spectroscopy
atomic absorption spectroscopy and mass spectroscopy
 
Information about the ionization spectroscopy technique
Information about the ionization spectroscopy techniqueInformation about the ionization spectroscopy technique
Information about the ionization spectroscopy technique
 
Liquid chromatography–mass spectrometry (LC-MS) BY P. RAVISANKAR
Liquid chromatography–mass spectrometry (LC-MS) BY P. RAVISANKARLiquid chromatography–mass spectrometry (LC-MS) BY P. RAVISANKAR
Liquid chromatography–mass spectrometry (LC-MS) BY P. RAVISANKAR
 
Bilal Ibrahim Dan-Iya_GS57595.pptx
Bilal Ibrahim Dan-Iya_GS57595.pptxBilal Ibrahim Dan-Iya_GS57595.pptx
Bilal Ibrahim Dan-Iya_GS57595.pptx
 
mass shruti
mass shrutimass shruti
mass shruti
 
liquid chromatography-mass spectrometry
liquid chromatography-mass spectrometryliquid chromatography-mass spectrometry
liquid chromatography-mass spectrometry
 

Más de Anubhav Gupta

Antitumor diabetics diuretics.pdf
Antitumor diabetics diuretics.pdfAntitumor diabetics diuretics.pdf
Antitumor diabetics diuretics.pdfAnubhav Gupta
 
antihyper tussive rheumatics.pdf
antihyper tussive rheumatics.pdfantihyper tussive rheumatics.pdf
antihyper tussive rheumatics.pdfAnubhav Gupta
 
Study of Crude Drugs acting on nervous system Part 3
Study of Crude Drugs acting on nervous system Part 3Study of Crude Drugs acting on nervous system Part 3
Study of Crude Drugs acting on nervous system Part 3Anubhav Gupta
 
Study of Crude Drugs Part 2
Study of Crude Drugs Part 2Study of Crude Drugs Part 2
Study of Crude Drugs Part 2Anubhav Gupta
 
STUDY OF CRUDE DRUGS PART 1 LAXATIVES
STUDY OF CRUDE DRUGS PART 1 LAXATIVESSTUDY OF CRUDE DRUGS PART 1 LAXATIVES
STUDY OF CRUDE DRUGS PART 1 LAXATIVESAnubhav Gupta
 

Más de Anubhav Gupta (6)

Antitumor diabetics diuretics.pdf
Antitumor diabetics diuretics.pdfAntitumor diabetics diuretics.pdf
Antitumor diabetics diuretics.pdf
 
antihyper tussive rheumatics.pdf
antihyper tussive rheumatics.pdfantihyper tussive rheumatics.pdf
antihyper tussive rheumatics.pdf
 
Study of Crude Drugs acting on nervous system Part 3
Study of Crude Drugs acting on nervous system Part 3Study of Crude Drugs acting on nervous system Part 3
Study of Crude Drugs acting on nervous system Part 3
 
Study of Crude Drugs Part 2
Study of Crude Drugs Part 2Study of Crude Drugs Part 2
Study of Crude Drugs Part 2
 
STUDY OF CRUDE DRUGS PART 1 LAXATIVES
STUDY OF CRUDE DRUGS PART 1 LAXATIVESSTUDY OF CRUDE DRUGS PART 1 LAXATIVES
STUDY OF CRUDE DRUGS PART 1 LAXATIVES
 
Nmr spectroscopy
Nmr  spectroscopyNmr  spectroscopy
Nmr spectroscopy
 

Último

Davis plaque method.pptx recombinant DNA technology
Davis plaque method.pptx recombinant DNA technologyDavis plaque method.pptx recombinant DNA technology
Davis plaque method.pptx recombinant DNA technologycaarthichand2003
 
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptx
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptxLIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptx
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptxmalonesandreagweneth
 
Vision and reflection on Mining Software Repositories research in 2024
Vision and reflection on Mining Software Repositories research in 2024Vision and reflection on Mining Software Repositories research in 2024
Vision and reflection on Mining Software Repositories research in 2024AyushiRastogi48
 
Pests of jatropha_Bionomics_identification_Dr.UPR.pdf
Pests of jatropha_Bionomics_identification_Dr.UPR.pdfPests of jatropha_Bionomics_identification_Dr.UPR.pdf
Pests of jatropha_Bionomics_identification_Dr.UPR.pdfPirithiRaju
 
Pests of soyabean_Binomics_IdentificationDr.UPR.pdf
Pests of soyabean_Binomics_IdentificationDr.UPR.pdfPests of soyabean_Binomics_IdentificationDr.UPR.pdf
Pests of soyabean_Binomics_IdentificationDr.UPR.pdfPirithiRaju
 
Microphone- characteristics,carbon microphone, dynamic microphone.pptx
Microphone- characteristics,carbon microphone, dynamic microphone.pptxMicrophone- characteristics,carbon microphone, dynamic microphone.pptx
Microphone- characteristics,carbon microphone, dynamic microphone.pptxpriyankatabhane
 
Fertilization: Sperm and the egg—collectively called the gametes—fuse togethe...
Fertilization: Sperm and the egg—collectively called the gametes—fuse togethe...Fertilization: Sperm and the egg—collectively called the gametes—fuse togethe...
Fertilization: Sperm and the egg—collectively called the gametes—fuse togethe...D. B. S. College Kanpur
 
Behavioral Disorder: Schizophrenia & it's Case Study.pdf
Behavioral Disorder: Schizophrenia & it's Case Study.pdfBehavioral Disorder: Schizophrenia & it's Case Study.pdf
Behavioral Disorder: Schizophrenia & it's Case Study.pdfSELF-EXPLANATORY
 
Call Girls in Majnu Ka Tilla Delhi 🔝9711014705🔝 Genuine
Call Girls in Majnu Ka Tilla Delhi 🔝9711014705🔝 GenuineCall Girls in Majnu Ka Tilla Delhi 🔝9711014705🔝 Genuine
Call Girls in Majnu Ka Tilla Delhi 🔝9711014705🔝 Genuinethapagita
 
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdfPests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdfPirithiRaju
 
Harmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms PresentationHarmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms Presentationtahreemzahra82
 
GenBio2 - Lesson 1 - Introduction to Genetics.pptx
GenBio2 - Lesson 1 - Introduction to Genetics.pptxGenBio2 - Lesson 1 - Introduction to Genetics.pptx
GenBio2 - Lesson 1 - Introduction to Genetics.pptxBerniceCayabyab1
 
Microteaching on terms used in filtration .Pharmaceutical Engineering
Microteaching on terms used in filtration .Pharmaceutical EngineeringMicroteaching on terms used in filtration .Pharmaceutical Engineering
Microteaching on terms used in filtration .Pharmaceutical EngineeringPrajakta Shinde
 
Neurodevelopmental disorders according to the dsm 5 tr
Neurodevelopmental disorders according to the dsm 5 trNeurodevelopmental disorders according to the dsm 5 tr
Neurodevelopmental disorders according to the dsm 5 trssuser06f238
 
Citronella presentation SlideShare mani upadhyay
Citronella presentation SlideShare mani upadhyayCitronella presentation SlideShare mani upadhyay
Citronella presentation SlideShare mani upadhyayupadhyaymani499
 
Carbon Dioxide Capture and Storage (CSS)
Carbon Dioxide Capture and Storage (CSS)Carbon Dioxide Capture and Storage (CSS)
Carbon Dioxide Capture and Storage (CSS)Tamer Koksalan, PhD
 
Pests of Bengal gram_Identification_Dr.UPR.pdf
Pests of Bengal gram_Identification_Dr.UPR.pdfPests of Bengal gram_Identification_Dr.UPR.pdf
Pests of Bengal gram_Identification_Dr.UPR.pdfPirithiRaju
 
OECD bibliometric indicators: Selected highlights, April 2024
OECD bibliometric indicators: Selected highlights, April 2024OECD bibliometric indicators: Selected highlights, April 2024
OECD bibliometric indicators: Selected highlights, April 2024innovationoecd
 

Último (20)

Davis plaque method.pptx recombinant DNA technology
Davis plaque method.pptx recombinant DNA technologyDavis plaque method.pptx recombinant DNA technology
Davis plaque method.pptx recombinant DNA technology
 
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptx
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptxLIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptx
LIGHT-PHENOMENA-BY-CABUALDIONALDOPANOGANCADIENTE-CONDEZA (1).pptx
 
Hot Sexy call girls in Moti Nagar,🔝 9953056974 🔝 escort Service
Hot Sexy call girls in  Moti Nagar,🔝 9953056974 🔝 escort ServiceHot Sexy call girls in  Moti Nagar,🔝 9953056974 🔝 escort Service
Hot Sexy call girls in Moti Nagar,🔝 9953056974 🔝 escort Service
 
Vision and reflection on Mining Software Repositories research in 2024
Vision and reflection on Mining Software Repositories research in 2024Vision and reflection on Mining Software Repositories research in 2024
Vision and reflection on Mining Software Repositories research in 2024
 
Pests of jatropha_Bionomics_identification_Dr.UPR.pdf
Pests of jatropha_Bionomics_identification_Dr.UPR.pdfPests of jatropha_Bionomics_identification_Dr.UPR.pdf
Pests of jatropha_Bionomics_identification_Dr.UPR.pdf
 
Pests of soyabean_Binomics_IdentificationDr.UPR.pdf
Pests of soyabean_Binomics_IdentificationDr.UPR.pdfPests of soyabean_Binomics_IdentificationDr.UPR.pdf
Pests of soyabean_Binomics_IdentificationDr.UPR.pdf
 
Microphone- characteristics,carbon microphone, dynamic microphone.pptx
Microphone- characteristics,carbon microphone, dynamic microphone.pptxMicrophone- characteristics,carbon microphone, dynamic microphone.pptx
Microphone- characteristics,carbon microphone, dynamic microphone.pptx
 
Fertilization: Sperm and the egg—collectively called the gametes—fuse togethe...
Fertilization: Sperm and the egg—collectively called the gametes—fuse togethe...Fertilization: Sperm and the egg—collectively called the gametes—fuse togethe...
Fertilization: Sperm and the egg—collectively called the gametes—fuse togethe...
 
Behavioral Disorder: Schizophrenia & it's Case Study.pdf
Behavioral Disorder: Schizophrenia & it's Case Study.pdfBehavioral Disorder: Schizophrenia & it's Case Study.pdf
Behavioral Disorder: Schizophrenia & it's Case Study.pdf
 
Call Girls in Majnu Ka Tilla Delhi 🔝9711014705🔝 Genuine
Call Girls in Majnu Ka Tilla Delhi 🔝9711014705🔝 GenuineCall Girls in Majnu Ka Tilla Delhi 🔝9711014705🔝 Genuine
Call Girls in Majnu Ka Tilla Delhi 🔝9711014705🔝 Genuine
 
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdfPests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdf
 
Volatile Oils Pharmacognosy And Phytochemistry -I
Volatile Oils Pharmacognosy And Phytochemistry -IVolatile Oils Pharmacognosy And Phytochemistry -I
Volatile Oils Pharmacognosy And Phytochemistry -I
 
Harmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms PresentationHarmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms Presentation
 
GenBio2 - Lesson 1 - Introduction to Genetics.pptx
GenBio2 - Lesson 1 - Introduction to Genetics.pptxGenBio2 - Lesson 1 - Introduction to Genetics.pptx
GenBio2 - Lesson 1 - Introduction to Genetics.pptx
 
Microteaching on terms used in filtration .Pharmaceutical Engineering
Microteaching on terms used in filtration .Pharmaceutical EngineeringMicroteaching on terms used in filtration .Pharmaceutical Engineering
Microteaching on terms used in filtration .Pharmaceutical Engineering
 
Neurodevelopmental disorders according to the dsm 5 tr
Neurodevelopmental disorders according to the dsm 5 trNeurodevelopmental disorders according to the dsm 5 tr
Neurodevelopmental disorders according to the dsm 5 tr
 
Citronella presentation SlideShare mani upadhyay
Citronella presentation SlideShare mani upadhyayCitronella presentation SlideShare mani upadhyay
Citronella presentation SlideShare mani upadhyay
 
Carbon Dioxide Capture and Storage (CSS)
Carbon Dioxide Capture and Storage (CSS)Carbon Dioxide Capture and Storage (CSS)
Carbon Dioxide Capture and Storage (CSS)
 
Pests of Bengal gram_Identification_Dr.UPR.pdf
Pests of Bengal gram_Identification_Dr.UPR.pdfPests of Bengal gram_Identification_Dr.UPR.pdf
Pests of Bengal gram_Identification_Dr.UPR.pdf
 
OECD bibliometric indicators: Selected highlights, April 2024
OECD bibliometric indicators: Selected highlights, April 2024OECD bibliometric indicators: Selected highlights, April 2024
OECD bibliometric indicators: Selected highlights, April 2024
 

mass spectroscopy

  • 1. ANUBHAV GUPTA B. PHARM 4TH YEAR Institute of Pharmacy VBS Purvanchal University
  • 2. 1.INTRODUCTION ▪ Mass Spectrometry (MS) is a technique that helps to identify the amount and type of chemicals present in a sample by measuring the mass-to-charge ratio and abundance of gas-phase ions. ▪ A mass spectrometer works by generating charged molecules or molecular fragments either in a high vacuum or immediately prior to the sample entering the high-vacuum region. ▪ Mass spectrum is a plot of relative abundance against the ratio of mass/charge (m/e). ▪ These spectra are used to determine the elemental or isotopic signature of a sample, the masses of particles and of molecules, and to elucidate the chemical structures of molecules and other chemical compounds. ▪ Mass spectrometry provides a highly specific method for determining or confirming the identity or structure of drugs and raw materials used in their manufacture. ▪ Mass spectrometry in conjunction with either gas chromatography (GC–MS) or liquid chromatography (LC–MS) provides a method for characterising impurities in drugs and formulation excipients. 2 Anubhav Gupta
  • 3. 2.PRINCIPLE ▪ The mass spectroscopy works on principle that when charged molecules or molecular fragments are generated in a high-vacuum region, or immediately prior to a sample entering a high-vacuum region, using a variety of methods for ion production. ▪ The ions are generated in the gas phase so that they can then be manipulated by the application of either electric or magnetic fields to enable the determination of their molecular weights. 3 Anubhav Gupta
  • 4. 3.Working of Mass Spectrometry ▪ In a typical procedure, a sample, which may be solid, liquid, or gas, is ionized, for example by bombarding it with electrons. ▪ This may cause some of the sample’s molecules to break into charged fragments. These ions are then separated according to their mass-to-charge ratio, typically by accelerating them and subjecting them to an electric or magnetic field: ▪ Ions of the same mass-to-charge ratio will undergo the same amount of deflection. ▪ The ions are detected by a mechanism capable of detecting charged particles, such as an electron multiplier. ▪ Results are displayed as spectra of the relative abundance of detected ions as a function of the mass-to-charge ratio. ▪ The atoms or molecules in the sample can be identified by correlating known masses (e.g. an entire molecule) to the identified masses or through a characteristic fragmentation pattern. 4 Anubhav Gupta
  • 5. 4.FRAGMENTATION PATTERN ▪ Fragmentation is the dissociation of energetically unstable molecular ions formed from passing the molecules in the ionization chamber of a mass spectrometer. ▪ Fragmentation of the molecular ions takes place in following forms: 1. Homolytic cleavage. This type of fragmentation is promoted by the presence of a hetero atom such as oxygen, nitrogen or sulphur, and in molecules containing a hetero atom it often gives rise to the most abundant ion in the mass spectrum 2. Heterolytic cleavage. Heterolytic cleavage occurs less often as a predominant fragmentation mechanism in drug molecules since they usually contain a lot of heteroatoms to direct the cleavage. In such cleavage, the positive charge is carried by the carbon atom not by the heteroatom. 3. Retro Diels-Alders reaction. It occurs in compounds with ring systems. It involves the cleavage of two bonds of a cyclic system which results in the formation of two stable unsaturated fragment in which two new bonds are formed. 5 Anubhav Gupta
  • 6. 4.1.Fragmentation Pattern ALOCOHOL(3-Pentanol) ALDEHYDE(3-Phenyl 2-Propenal) ALKANE(Hexane) AMIDE(Methyl butyramide) 6 Anubhav Gupta
  • 7. 5.McLafferty Rearrangement ▪ The McLafferty rearrangement is a characteristic fragmentation of the molecular ion of a carbonyl compound containing at least one gamma hydrogen, e.g., Migration of -hydrogen followed by -bond cleavage and elimination of ethylene or substituted ethylene neutral molecule. ▪ The McLafferty rearrangement, is relatively uncommon in drug molecules; it is most often encountered in fatty acid esters. ▪ This rearrangement can occur in carbonyl compounds e.g. amides, ketones and acids. 7 Anubhav Gupta
  • 8. 6.IONIZATION TECHNIQUES ▪ There are many types of ionization techniques that are used in mass spectrometry. Some of them are: 1.Electron Impact ionization (EI) 2.Fast Atom Bombardment (FAB) 3.Electrospray ionization (ESI) 4.Atmospheric Pressure Chemical Ionization (APCI) 5.Matrix Assisted Laser Desorption Ionization (MALDI) ▪ APCI is considered best than EI because APCI forms a protonated molecule and is completely compatible with Liquid Chromatography (LC), while EI is more likely to fragment the ion leading to a possible more ambiguous identification of the molecule weight and is incompatible with LC. ▪ MALDI along with ESI allows for ionization and measurement of large molecular weight. ▪ ESI has an advantage in its easy compatibility with LC. While MALDI has advantages for imaging mass spectrometry. 8 Anubhav Gupta
  • 9. 9 6.1. Electron Impact Ionization(EI) ▪ EI is still used in conjunction with sample introduction either via a direct heated probe or via gas chromatography (GC): (i) The sample is introduced into the instrument source by heating it on the end of a probe until it evaporates, assisted by the high vacuum within the instrument or via a capillary GC column. (ii) Once in the vapour phase, the analyte is bombarded with the electrons produced by a rhenium or tungsten filament, which are accelerated towards a positive target with an energy of 70 eV. The analyte is introduced between the filament and the target, and the electrons cause ionisation as follows: M þ e ! Mþ: þ 2 e. (iii) Since the electrons used are of much higher energy than the strength of the bonds within the analyte (4–7 eV), extensive fragmentation of the analyte usually occurs. (iv) The molecule and its fragments are pushed out of the source by a repeller plate which has the same charge as the ions generated. Anubhav Gupta
  • 10. 10 6.2. Fast Atom Bombardment(FAB) ▪ FAB is a technique that was popular in the 80's to early 90's because it was the first technique that allowed ionization of non-volatile compounds that could be done simply. ▪ It was done by bombarding a sample in a vacuum with a beam of atoms, typically Ar or Xe, accelerated to Kilovolt energies. The sample was typically mixed in a matrix. ▪ The two most common matrixes were glycerol and 3 Nitro-benzoic acid. The matrix allowed the sample to refresh itself. ▪ The ions formed by FAB were adducts to the molecule, where the adducts could be protons, sodium ions, potassium ions or ammonium ions. ▪ A variation of FAB was replacement of the atom beam with a beam of ions, typically cesium ions, which was called secondary ion mass spectrometry (SIMS). SIMS spectra were typically identical to FAB spectra and the terms became interchangeable. Anubhav Gupta
  • 11. 11 6.3. Electronspray Ionization(ESI) ▪ The electronspray is created by putting a high voltage on a flow of liquid at atmospheric pressure, sometimes this is assisted by a concurrent flow of gas. ▪ The created spray is directed to an opening in the vacuum system of the mass spectrometer, where the droplets are de-solvated by a combination of heat, vacuum and acceleration into gas by voltages. ▪ The ions are ejected in droplets and accelerated into the mass analyzer by voltages. For larger molecules, the ions may contain multiple charges, allowing the detection of very large molecules on analyzers that have limited mass to charge (m/Z)) ratio ranges. Because of the natural use of a flowing liquid, it is easily adapted to liquid chromatography (LC). ▪ There are many other techniques that are variations of electrospray, for example, nanospray, picospray desorption electrospray ionization (DESI). Anubhav Gupta
  • 12. 12 6.4. Atmospheric Pressure Chemical Ionization (APCI) ▪ Atmospheric pressure chemical ionisation (APCI) is closely related to ESI and instruments carrying out ESI can be readily switched to operate in APCI mode. ▪ In APCI mode the eluent from the HPLC does not pass through a charged needle before entering the mass spectrometer source but via a heated tube so that it forms an aerosol. Upon exiting the heated tube an electric discharge is passed through the aerosol generating reactive species such as H3Oþ and N2þ, which promote the ionisation of the analytes. ▪ The technique has never achieved the level of popularity of ESI since most drug molecules can be ionised under ESI conditions; however, APCI can be employed for the analysis of drug molecules of low polarity that do not ionise efficiently under ESI conditions. Anubhav Gupta
  • 13. 13 6.5. Matrix Assisted Laser Desorption Ionization (MALDI) ▪ MALDI uses a nitrogen laser to promote ionisation of molecules prior to ion separation in a mass spectrometer. It is usually combined with time of flight (TOF) separation of the ions generated. ▪ In order for the sample to be ionised it has to be dissolved in a matrix that absorbs UV radiation at around the wavelength (337 nm) produced by the laser. A simple example of a matrix is dihydroxybenzoic acid and there are a number of similar aromatic compounds which are used to promote ionisation of different classes of molecules. ▪ The sample solution is mixed with matrix solution on a metal plate and allowed to dry prior to being introduced into the instrument. The laser is then directed at the target plate to promote ionisation. Anubhav Gupta
  • 14. 14 7. INSTRUMENTATION Mass spectrometer has following components: A. Sample Inlet, through HPLC, GC, Syringe etc. B. Ionization, can be achieved by : 1.Electron Impact ionization (EI) 2.Fast Atom Bombardment (FAB) 3.Electrospray ionization (ESI) 4.Atmospheric Pressure Chemical Ionization (APCI) 5.Matrix Assisted Laser Desorption Ionization (MALDI) C. Acceleration & Deflection D. Analyser, can be achieved by analysers like : 1.Magnetic sector mass analyser 2.Double focussing analyser 3.Quadrupole mass analyser 4.Time of Flight analyser (TOF) 5.Ion trap analyser 6.Ion cyclotron analyser E. Detector Anubhav Gupta
  • 15. 15 7.1. Sample Inlet ▪ Sample stored in large reservoir from which molecules reaches ionization chamber at low pressure in steady stream by a pinhole called “Molecular leak”. ▪ Molecular leak–It is pin-hole restriction (0.01 to 0.05mm diameter) and made up of gold foil. It is used for metering the sample to ionization chamber. 7.2. Ionization ▪ Atoms are ionized by knocking one or more electrons off to give positive ions by bombardment with a stream of electrons. Most of the positive ions formed will carry charge of +1. 7.3.Acceleration & Deflection ▪ Ions are accelerated so that they all have same kinetic energy. ▪ Positive ions pass through 3 slits with voltage in decreasing order. ▪ Middle slit carries intermediate and finals at zero volts. ▪ Ions are deflected by a magnetic field due to difference in their masses. ▪ The lighter the mass, more they are deflected. ▪ It also depends upon the no. of +ve charge an ion is carrying; the more +ve charge, more it will be deflected. Anubhav Gupta
  • 16. 16 7.4. Analyser ▪ A mass analyser is a device that can separate atoms and molecules according to their mass. ▪ The five main characteristics for measuring the performance of a mass analyser are 1) The mass range limit or dynamic range 2) The analysis speed [u (m)S-1] 3) The transmission = No. of ion reaching the ions/No. of ions entering mass analyzer 4) The mass accuracy 5) The resolution. ▪ Mass analysers used are: 1.Magnetic sector mass analyser 2.Double focussing analyser 3.Quadrupole mass analyser 4.Time of Flight analyser (TOF) 5.Ion trap analyser 6.Ion cyclotron analyser Anubhav Gupta
  • 17. 17 7.4.1. Quadrupole Mass Analyzer ▪ A cheaper and more sensitive mass spectrometer than a magnetic sector instrument is based on the quadrupole analyser. ▪ It uses two electric fields applied at right angles to each other, rather than a magnetic field, to separate ions according to their m/z ratios. One of the fields used is DC and the other oscillates at radiofrequency. Disadvantage ▪ Limited resolution ▪ Peak heights variable as a function of mass (mass discrimination). ▪ Peak height vs. mass response must be 'tuned'. ▪ Not well suited for pulsed ionization methods. ▪ Low-energy collision-induced dissociation (CID) MS/MS spectra in triple quadrupole and hybrid mass spectrometers depend strongly on energy, collision gas, pressure, and other factors. Anubhav Gupta
  • 18. 18 7.4.2. Time of flight (TOF) analyser ▪ The basis of the separation is that smaller ions move more quickly than larger ions and thus reach the detector first. ▪ The technique relies on gating the signal from the ion source, so that one packet of ions is allowed time to reach the detector before the next set is ejected from the ion source, so that there is no overlap between ion packets. The ions leaving the ion source have different kinetic energies and this compromises the mass resolution, indeed early instruments gave very wide peaks for a single mass. ▪ The problem of mass resolution was resolved by using a device called a reflectron, which opposes the direction that the ions are moving by sending them back in the opposite direction. ▪ The faster moving ions penetrate deeper into the reflectron and thus a lag effect is produced for faster moving ions so that they are focused with the slower moving ions which do not penetrate as far into the reflectron. Pushing the ions back in the opposite direction also increases the length of the flight tube, increasing instrument resolution. Anubhav Gupta
  • 19. 19 7.4.3. Magnetic Sector Mass Analyzer ▪ In a magnetic sector instrument the ions generated are pushed out of the source by a repeller potential of same charge as the ion itself (most often positive). ▪ They are then accelerated in an electric field of ca 3–8 kV and travel through an electrostatic field region so that they are forced to fall into a narrow range of kinetic energies prior to entering the field of a circular magnet. ▪ They then adopt a flight path through the magnetic field depending on their charge to mass (m/z) ratio; the large ions are deflected less by the magnetic field: 𝒎 𝒛 = 𝑯𝟐𝒓𝟐 𝟐𝑽 where H is the magnetic field strength, r is the radius of the circular path in which the ion travels, and V is the accelerating voltage. Anubhav Gupta
  • 20. 20 Comparison of various analyzers Anubhav Gupta
  • 21. 21 7.5. Detectors ▪ The ion collection system measures the relative abundance of ion fragments of each mass. ▪ Several types of detectors are available for mass spectrometers. The detector used for most routine experiments is the electron multiplier. ▪ Another type of detector is photographic plates coated with a silver bromide emulsion, it is sensitive to energetic ions. A photographic plate can give a higher resolution than an electrical detector. Faraday Cup: ▪ It consists of a hollow conducting electrode connected to ground via a high resistance. ▪ The ions hitting the collector cause a flow of electrons from ground to the resistor. ▪ The resulting potential drop across the resistor is amplified. A plate held at about -80 V in front of the collector, prevents any ejected secondary electrons from escaping and causing an anomalous reading. Photographic Plates: ▪ This detector system is most sensitive than any other detector because the photoplate integrates the ion signal over a period of time. ▪ The photoplate are processed by the usual photographic techniques and read with the aid of densitometer. Anubhav Gupta
  • 22. 22 Electron Multipliers: ▪ It is most common means of detecting ions. It is made up of a series (12 to 24) of aluminum oxide dynodes maintained at ever increasing potentials. ▪ Ions strike the first dynode surface causing an emission of electrons. These electrons are then attracted to the next dynode held at a higher potential and therefore more secondary electrons are generated. ▪ Ultimately, as numerous dynodes are involved, a cascade of electrons is formed that reult in an overall current gain on the order of one million or higher. ▪ The high energy dynode(HED) uses an electrostatic field to increase the velocity of the ions and serves to increase signal intensity and therefore sensitivity. Faraday Cup Photoplate Electron Multiplier Anubhav Gupta
  • 23. 23 8.APPLICATIONS ▪ Mass spectrometry provides a highly specific method for determining or confirming the identity or structure of drugs and raw materials used in their manufacture. ▪ Mass spectrometry in conjunction with either gas chromatography (GC–MS) or liquid chromatography (LC–MS) provides a method for characterising impurities in drugs and formulation excipients. ▪ Pharmaceutical Analysis Bioavailability studies Drug metabolism studies Characterization of potential drugs Drug degradation product analysis Identifying drug targets ▪ Biomolecular characterization Characterization of Proteins an Peptides Oligonucleotides. ▪ Environmental Analysis ▪ Forensic Analysis Anubhav Gupta